SIGNA™ Architect

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: SIGNA™ Architect
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
SIGNA™ Architect905684MR3TSIGNA ARCHITECT

Issue

GE Healthcare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels may exceed 99db, the limit established by international electrotechnical commission (IEC 60601-2-33) when using hearing protection with a noise reduction rating (NRR) of 29db, as currently required in the operator manual. Prolonged periods of exposure to elevated acoustic noise could potentially lead to hearing impairment.

Recall start date: October 31, 2024