Health Products

COBE 2991 Disposable Sets

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: COBE 2991 Disposable Sets
Issue: Medical devices - Performance
What to do: Instructions to be provided by manufacturer on how to mitigate risk and not use affected devices.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
COBE 2991 Blood Processing SetMore than 10 numbers, contact manufacturer.90819
Triple Processing Set10G15203
07G15207
02G15222
90901

Issue

Three (3) separate failure modes have been identified that may lead to procedural interruptions, leaks, and subsequent loss of product during use of select COBE 2991 disposable sets. These failure modes include: 1) an improperly bonded spike receptor on the processing bag, 2) a twisted stem tube due to a seized ceramic bearing, and 3) a small hole (a.k.a. pinhole) on the stem tube below the rotating seal.

Recall start date: October 21, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime