Health Products

Plum Infusion Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use, Health products - Medical devices - General and plastic surgery
Product: Plum Infusion Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Plum A+ Infusion SystemAll lots.20792-XX-YY
12391-XX-YY
Plum A+ Hyperbaric Infusion SystemAll lots.11005
Plum 360 Infusion SystemAll lots.30010-XX-YY
Plum A+3 Infusion SystemAll lots.20678-XX-YY
20679-XX-YY
12618-XX-YY

Issue

ICU Medical has received reports of counterfeit CSB batteries being used with Plum infusion systems. While these counterfeit batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum infusion systems. Preliminary reports suggest that these counterfeit batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. Do not use spare parts that are not authorized by ICU Medical. Please ensure that any entity providing service or repair activities for your plum infusion systems uses only parts authorized or directly provided by ICU Medical.

Recall start date: October 21, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime