Health Products

Cartiva Synthetic Cartilage Implant

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Cartiva Synthetic Cartilage Implant
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Cartiva Synthetic Cartilage ImplantAll lots.CAR-06
CAR-10
CAR-08
CMC-08

Issue

Stryker has become aware of recently published data and post market reports indicating that patients implanted with Cartiva SCI may experience a higher-than- expected occurrence rate when compared to data submitted in the 2016 PMA of the following documented hazards: revision, removal, implant subsidence, displacement, pain, nerve damage or fragmentation. Cartiva SCI devices have been observed in some cases to be revised/removed at higher rates than previously observed in the initial Cartiva SCI premarket and post-approval studies.

Recall start date: October 31, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime