Atellica CI System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Atellica CI System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Atellica CI SystemIRC009062411
IRC007512342
IRC009012411
10947347

Issue

Siemens Healthineers has confirmed through the internal investigation of customer complaints that the IMT diluent volume remaining (% remaining) does not decrease as expected on the Atellica® CI analyzer. The IMT module runs a standard a refresh sequence before processing samples if at least two minutes of inactivity have passed. IMT diluent is used by this sequence, however, the calculated diluent volume is not accurate, potentially leading to the IMT diluent being empty while still displaying that volume is remaining. In this case sodium (Na), potassium (K) and chloride (Cl) test results may be falsely elevated. quality control materials demonstrate the same behavior.

Recall start date: November 6, 2024