· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| BD BBL Sensi Disc Ampicillin 2 Μg (AM-2) | 3234190 3184064 2339360 3010977 | 231263 |
BD identified through potency test, as part of special stability test requests to monitor Ampicillin AM-2, unsatisfactory results of 65% at 18 months. The potency acceptance criteria is 80-140% and product shelf life is 24 months. A retest with control batch and extended sampling confirmed the low potency results.
Recall Start Date: November 6, 2024