Health Products

TDHistoCyto Software

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: TDHistoCyto Software
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
TDHistoCyto SoftwareSoftware version V13.21.BTD-TDH-01

Issue

Inappropriate behaviour was noted on a TDHC user site, specifically with regard to send/receipt by an external host system. In some rare cases, analysis reports were included in the hospital information system (HIS) with results from another patient. This recall applies only to systems with an INST0031 connection.

Recall Start Date: August 28, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime