· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Proteus XR/a General Purpose Radiographic System | More than 10 numbers, contact manufacturer. | 2259988 2268970 2259973-3 |
GE Healthcare has become aware that steel cables in the wall stand for certain Proteus XR/a systems serviced by GE Healthcare have not been replaced at the intervals detailed in the preventative maintenance action list (GEHC service manual 2273022-100).
Recall Start Date: November 4, 2024