Health ProductsType I recall

Access Intact PTH

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Access Intact PTH
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Access Intact PTH339071A16972

Issue

Beckman Coulter has determined that 91 Access Intact PTH reagent packs (lot 339071) were filled with an incorrect volume of reagent in well 0.

Access Intact PTH patient and QC samples with a reagent pack that has less volume than expected may: - Recover up to 30% lower than the expected results for routine PTH, with a pack imprecision of up to 10%. - Recover up to 40% lower than the expected results for intraoperative PTH, with a pack imprecision of up to 15%. - Produce a zero or near-zero result.

Falsely decreased Intact PTH results could lead to a delay in diagnosis, additional diagnostic testing, and/or inappropriate treatment.

Recall Start Date: November 25, 2024