· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Access hsTnI | All lots. | B52699 |
In remote circumstances, if a sample containing high cardiac troponin (CTNI) >55,000 pg/ml is tested using an access assay other than Access hsTnI, CTNI carryover may occur if the next test performed, using the same probe on that instrument, is Access hsTnI on a different sample.
Recall start date: November 18, 2024