Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Access hsTnI
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Access hsTnIAll lots.B52699

Issue

In remote circumstances, if a sample containing high cardiac troponin (CTNI) >55,000 pg/ml is tested using an access assay other than Access hsTnI, CTNI carryover may occur if the next test performed, using the same probe on that instrument, is Access hsTnI on a different sample.