Health Products

EGM - Essential Gas Module

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: EGM - Essential Gas Module
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
EGM - Essential Gas ModuleNot applicable.866173

Issue

Philips has identified a potential safety issue for the 866173 IntelliVue G7m anesthesia gas modules, which could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, inop (inoperable) alarm. It was anticipated that the expected operating lifetime of the pump was at least 10000 hours. However, based on testing, the average operating lifetime was found to be 4500 hours. The pump may fail prematurely when there is friction caused by an abrasion.

Recall restart date: February 14, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime