· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| BD FACSLyric Flow Cytometer | More than 10 numbers, contact manufacturer. | 659180 651164 663029 651165 |
BD has identified through complaints an increase in field failures / replacements of the power supply module on BD FACSLyric Flow Cytometers. The capacitor on the power supply can split which has the potential for the instruments to fail to power on and/or stay on.
Recall Start Date: November 20, 2024