· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Zeus Infinity Empowered | More than 10 numbers, contact manufacturer. | MK04000 |
The Zeus IE features two gas measurement modules, a patient gas analyzer, and a system gas analyzer. Both contain a pump which is sampling the gas to be analyzed. This sampling pump has shown an increased failure rate, which might lead to different effects during use of the Zeus IE. The effects depend on the therapy settings.
Recall Start Date: July 31, 2024