· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| ACCOLADE™ Pacemaker | More than 10 numbers, contact manufacturer. | L301 L300 L321 |
| VISIONIST™ CRT-P | More than 10 numbers, contact manufacturer. | U225 |
| ACCOLADE™ MRI Pacemaker | More than 10 numbers, contact manufacturer. | L331 L310 L311 |
| VISIONIST™ X4 Crt-P | More than 10 numbers, contact manufacturer. | U228 |
Boston Scientific is initiating an advisory to provide information about a subset of pacemakers from the ACCOLADE™ family and VISIONIST™ CRT-Ps with an increased potential to initiate safety mode during telemetry or, in rare instances, other normal, higher-power operations due to latent, high battery impedance when the device reaches approximately four (4) years or less of remaining battery longevity.
Recall start date: December 12, 2024