Health Products

Maquet Hemoconcentrator Set With Softline Coating

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Hemoconcentrator Set With Softline Coating
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Hemoconcentrator Set With Softline Coating3000382838
701052273
3000361271
701074294
3000407060
701035333
3000393741
BO-H 48773

Issue

An issue with eight technical drawings, specifically the absence of the bonding symbol for certain connections on the tubing set was found. It is important to note that the specified bonding procedure, which involved the use of cyclohexanone, was not carried out. While the absence of the bonding on the connectors may not be immediately apparent, it could potentially lead to leakage in the affected connections.

Recall start date: December 9. 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime