Health ProductsType I recall

Neo-Tee T-Piece Resuscitator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Neo-Tee T-Piece Resuscitator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Neo-Tee T-Piece Resuscitator242645080910-50809

Issue

To address the potential for performance specifications not being met. The device may not reach the high and the low ends of the peak inspiration pressure (PIP) and positive end expiratory pressure (PEEP) ranges as a result of an undersized spring in the controller.

Recall start date: December 5, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime