QAngio XA 3D

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: QAngio XA 3D
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
QAngio XA 3DNot applicable2.2.2.4

Issue

For certain Siemens XA images, the isocenter calibration value is calculated incorrectly due to an incorrect value for distance source to isocenter that is saved by Siemens in the standard dicom attribute (0018,1111). The correct value is stored in a Siemens private dicom attribute (0021, xx17, "Siemens sms-ax acq 1.0") in the dicom data. Incorrect calibration can lead to errors in the measurements provided by qAngio XA 3D.

Recall start date: December 16, 2024