Health Products

Azurion and Allura Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Azurion and Allura Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Azurion 7 M20Not applicable.722 224
722 079
Allura Xper Fd20 Or Table (Product Of Alluraclarity Family)Not applicable.722035
Allura Xper Fd20 (Product Of Alluraclarity Family)Not applicable.722028

Issue

Philips has identified that due to the forces applied during the movement of the c-arc of the Azurion and Allura systems, the bolts supporting the FlexMove carriage may become loose and/or break, and cracks may appear in the FlexMove carriage. If FlexMove rail fixation bolts become loose or broken, or cracks appear in the carriage, the following issues may occur: • transversal movements of the c-arc stop due to false collision detections because of additional friction. • manual movement of the FlexMove carriage not possible due to added friction. • abnormal noise during transversal movement of the c-arc. • unstable c-arc suspension. • fall of the c-arc assembly (1,500 kg), if all bolts in the x-axis would break/get loose. • drop of the c-arc assembly (up to 10cm if c- arc is at one side of the rail, up to 5 cm if c-arc is in the center of the rails, and 1.5 cm if the c-arc is in the anterior/posterior position), if all bolts in the y-axis would break/get loose.

Recall start date: December 12, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime