Health Products

IP DAB Chromogen Kit 80 mL

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Pathology
Product: IP DAB Chromogen Kit 80 mL
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
IP DAB Chromogen Kit 80 mL100424
112823
071923
041624
091923
IPK 5010 G80

Issue

Biocare Medical initially became aware of an issue where certain kits containing the DAB Chromogen Buffer Lots were not performing as expected, producing reduced or weak to no staining. The initial recall in Sept 2024 did not include product distributed in Canada. Following further investigation, Biocare is expanding the scope of the recall to include additional kits containing the identified lots of DAB Buffer, including product distributed in Canada. Recall Start Date: December 19, 2024