Health Products

Nobel Biocare N1 TiUltra TCC NP

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Dental
Product: Nobel Biocare N1 TiUltra TCC NP
Issue: Medical devices - Sterility
What to do: ​​​​​​Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Nobel Biocare N1 TiUltra TCC NP12254213300863

Issue

A manufacturing error in one of the supplied packaging components may result in the presence of a sharp edge. This sharp edge could puncture the blister packaging and breach the sterile barrier. Recall Start Date: December 23, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime