Health Products

VACUETTE® Sample Transfer Unit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: VACUETTE® Sample Transfer Unit
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
VACUETTE® Sample Transfer UnitB240749R
B2407379
B24083NM
B24083AK
B240837T
B24063KJ
450218

Issue

Reports to Greiner indicate several lots of the sample transfer units (stu) may leak due to the protective rubber sleeve bunching up and not retracting when using the device. The non reset of the rubber sleeve can depend on the device age and can be up to 23%. The sample transfer unit (stu) is a custom-made, single-use tube holder, used solely in plasma centers, to transfer a small amount of plasma from the bottle to the tube. Once transferred in the tube, plasma samples are tested to assure quality standards/specifications of the bottled plasma are met prior to sale. no patient treatment is performed at plasma centers. Since there is no venous pressure and personal protective equipment (PPE) is required when handling plasma samples, the risk to the user is minimal. Any possible leakage does not impact the device's capability to collect plasma sample nor the quality of the plasma sample. As a result, the potential effects of the detected error are viewed as low risk. Nevertheless, the leakage could still pose a minimal risk to the user, specifically if appropriate PPE is not worn. Therefore, the concerned lots of the device should not be utilized. Recall start date: November 4, 2024