Health Products

Zeus® Infinity® Empowered

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Zeus® Infinity® Empowered
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Zeus® Infinity® EmpoweredMore than 10 numbers, contact manufacturer.MK04000

Issue

In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check. Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement. Recall start date: December 23, 2024