Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Digital Telemetry System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Digital Telemetry SystemMore than 10 numbers, contact manufacturer.96280

Issue

XTR systems with software versions 1.4.0 and 1.4.1 may experience an unsolicited system shut down and restart. While the XTR system is running and telemetry devices are being monitored from xhibit, the system may shut down, which will cause a temporary loss of monitoring. During this time, telemetry zones on the xhibit will state "offline" and monitoring will be lost until the system has restarted and comes back online. Recall start date: December 24, 2024