Vitros® 5600, 3600 and XT Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Vitros® 5600, 3600 and XT Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Vitros® 5600 Integrated SystemNot applicable.680 2413
680 2915
Vitros® 3600 Immunodiagnostic System-InstrumentNot applicable.680 2914
680 2783
Vitros® XT 7600 Integrated SystemNot applicable.6272222
684 4461

Issue

During internal testing, QuidelOrtho™ identified that microwell assay results have the potential to be reported without the reagent expired (re) result code, even when the signal reagent (sr) pack is past the expiry date. This can occur when an sr pack exceeds the on-board expiry and/or the shelf expiration date. This issue can affect any microwell assay regardless of whether or not the system is configured to use expired reagents. Recall start date: December 24, 2024