ETHIBOND EXCEL® , ETHIBOND®, PERMAHAND®, PROLENE® sutures

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: ETHIBOND EXCEL® , ETHIBOND®, PERMAHAND®, PROLENE® sutures
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
ETHIBOND® Extra Polyester Suture103H4BX425H
PROLENE® Suture - Taper Point103E228706H
PERMAHAND® Non Absorbable Silk Suture10386S
103DBT
623H
K872H
ETHIBOND EXCEL® Polyester Suture103ELKX872H

Issue

An open seal defect was detected during internal rework activities. Investigation determined that the preliminary cause of issue was a machine failure, leading to the zipper tray to be incorrectly positioned in the copolymer pre-formed cavity and subsequent tray misalignment with the front seal area in the sealing process. This ultimately resulted in possibility of open seal condition and/or narrow seal defects in the subsequent units. Recall start date: December 30, 2024