· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| AK 98, 115V, Self-Care | All lots. | 115251 |
| AK 98 V2 115v Bio Self-Care | All lots. | 955407 |
| AK 96, 115 V, Self-Care | All lots. | 110661 |
| AK 98v3 115v Self-Care | All lots. | 955606 |
| AK 96, 115 V, Bio | 20072 19489 | 110373 |
Baxter Healthcare Corporation (Baxter) is issuing an urgent medical device correction for the AK 96 and AK 98 dialysis machines with the self-care configuration listed below due to the potential for hemolysis caused by the arterial dialyzer line kinking from the position of the remote panel arm. Recall start date: January 16, 2025