Health Products

Essenz HLM

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Essenz HLM
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Essenz HLM49PK00427
49PK00012
49PK00047
49PK00432
49PK00048
49PK00426
49PK00430
49PK00045
49PK00429
49PK00433
49PK00424
49PK00419
49-00-10

Issue

LivaNova has received reports of the Cockpit Graphical User Interface (GUI) of the Essenz HLM with SW HLM.1.5 activating a self-resetting safety feature, with the operation of the GUI being restored during use. During the reported events, the touchscreen display has temporarily gone dark (black/grey) and reset back to the home screen after approximately 10-15 seconds. Recall Start Date: October 18, 2024