Health Products

MEDRAD® Centargo Injection System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: MEDRAD® Centargo Injection System
Issue: Medical devices - Performance
What to do: Customers will be notified with instructions to follow.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
MEDRAD® Centargo Injection SystemContact the manufacturer.CENT-SYS-BAT

Issue

Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low. Recall start date: December 13, 2024