VNS Therapy SenTiva® Duo Generator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Neurology
Product: VNS Therapy SenTiva® Duo Generator
Issue: Medical devices - Performance
What to do: Customers to be notified. Instructions to follow on response.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
VNS Therapy SenTiva® Duo GeneratorMore than 10 numbers, contact manufacturer.1000-D

Issue

Complaints have been received on the SenTiva Generator reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events. Recall start date: January 10, 2025