Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: DLP Antegrade Aortic Root Cardioplegia Cannulae
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
DLP Antegrade Aortic Root Cardioplegia CannulaeMore than 10 numbers, contact manufacturer.11012
11014
11014L
21014
11012L
21012

Issue

During the manufacturing process, unexpected loose plastic material in the male luer used in the aortic root cannula was identified. The potential harms when identified prior to use is procedure delay while another cannula is located. If this is not identified prior to use, and the clinician uses the cannula, the potential harm is stroke (reversible and irreversible). Recall Start Date: February 5, 2025