· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| MiniMed 670g | All lots. | MMT-1782K MMT-1782 MMT-1782KRP MMT-1782CRP |
| MiniMed 670g Kit | All lots. | MMT-1762KCN MMT-1762CRP |
| MiniMed 770g | All lots. | MMT-1881 MMT-1891CN |
| MiniMed 630g | All lots. | MMT-1714K MMT-1714KRP MMT-1714 |
| Paradigm 554 Insulin Pump | All lots. | MMT-554 |
| MiniMed 780g Pump | All lots. | MMT-1885 |
| MiniMed 780g Kit | All lots. | MMT-1895 |
| Paradigm 754 Insulin Pump | All lots. | MMT-754 |
When air pressure decreases (e.g., during flight takeoff), more insulin may be released than expected. Additionally, unintended insulin may be released even if the pump's delivery is suspended or programmed to zero units per hour. When air pressure increases (e.g., during flight landing), less insulin may be released than expected. Recall Start Date: January 28, 2025