MiniMed and Paradigm Insulin Pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: MiniMed and Paradigm Insulin Pumps
Issue: Medical devices - Performance
What to do: ​​​​​​Contact the manufacturer if you require additional information.
Audience: General public
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
MiniMed 670gAll lots.MMT-1782K
MMT-1782
MMT-1782KRP
MMT-1782CRP
MiniMed 670g KitAll lots.MMT-1762KCN
MMT-1762CRP
MiniMed 770gAll lots.MMT-1881
MMT-1891CN
MiniMed 630gAll lots.MMT-1714K
MMT-1714KRP
MMT-1714
Paradigm 554 Insulin PumpAll lots.MMT-554
MiniMed 780g PumpAll lots.MMT-1885
MiniMed 780g KitAll lots.MMT-1895
Paradigm 754 Insulin PumpAll lots.MMT-754

Issue

When air pressure decreases (e.g., during flight takeoff), more insulin may be released than expected. Additionally, unintended insulin may be released even if the pump's delivery is suspended or programmed to zero units per hour. When air pressure increases (e.g., during flight landing), less insulin may be released than expected. Recall Start Date: January 28, 2025