Access 2 Immunoassay Analyzers

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Access 2 Immunoassay Analyzers
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Access System Access 2 Analyzer - Class 2More than 10 numbers, contact manufacturer.81600N
Access Immunoassay System - Analyzer Class 3More than 10 numbers, contact manufacturer.81600N
Access2 Immunoassay SystemMore than 10 numbers, contact manufacturer.81600N

Issue

Beckman Coulter has determined that the sample motor on the affected Access 2 Immunoassay analyzers may have an improper alignment to the sample wheel that may result in irregular motion as the sample carousel turns. The irregular motion has the potential to cause damage to the sample wheel motor and/or the sample wheel index sensor over time and can cause an increase in "sample carousel motion error" event messages. After a "sample carousel motion error" occurs, if the analyzer cannot recover from the motion error upon reinitialization, there may be a delay in reporting patient test results to the physician. Recall start date: February 3, 2025