· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Assurity Pulse Generator | More than 10 numbers, contact manufacturer. | PM2240 |
| Assurity MRI Pacemaker | More than 10 numbers, contact manufacturer. | PM1272 PM2272 |
| Endurity Pulse Generator | More than 10 numbers, contact manufacturer. | PM1160 |
| Endurity MRI Pacemaker | More than 10 numbers, contact manufacturer. | PM1172 PM2172 |
Abbott is informing clinicians of the potential for a device malfunction affecting a subset of Assurity™ and Endurity™ pacemakers. This issue may result in incomplete mixing of epoxy during manufacturing and, with time, may permit moisture ingress into the pulse generator header, introducing a risk of interrupting device functionality. Affected devices were manufactured between August 2019 and June 2020. Recall start date: February 18, 2025