Health Products

Assurity and Endurity Pacemakers

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Assurity and Endurity Pacemakers
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Assurity Pulse GeneratorMore than 10 numbers, contact manufacturer.PM2240
Assurity MRI PacemakerMore than 10 numbers, contact manufacturer.PM1272
PM2272
Endurity Pulse GeneratorMore than 10 numbers, contact manufacturer.PM1160
Endurity MRI PacemakerMore than 10 numbers, contact manufacturer.PM1172
PM2172

Issue

Abbott is informing clinicians of the potential for a device malfunction affecting a subset of Assurity™ and Endurity™ pacemakers. This issue may result in incomplete mixing of epoxy during manufacturing and, with time, may permit moisture ingress into the pulse generator header, introducing a risk of interrupting device functionality. Affected devices were manufactured between August 2019 and June 2020. Recall start date: February 18, 2025