Health Products

Novasight Hybrid Catheter

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Novasight Hybrid Catheter
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Novasight Hybrid Catheter240502TA-06-0001

Issue

A portion of the catheter was detached and left in the patient's body during a clinical case in USA on Nov.21, 2024. The portion was retrieved by the physicians, and there was no patient injury reported. Health Canada FRN was submitted on Mar.8, 2024 following initiation of a voluntary recall in USA. As there is a probability that use of or exposure to the affected catheter lots may cause adverse health consequences, Conavi Medical Inc. proposes a voluntary recall of Novasight Hybrid Catheters in Canadian market starting March 12, 2025. Recall start date: Marc 12, 2025

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime