DxC 500 AU Clinical Chemistry Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: DxC 500 AU Clinical Chemistry Analyzer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
DxC 500 AU Clinical Chemistry Analyzer2024070118C63520
C63519

Issue

Beckman Coulter has determined that the DxC 500 AU Clinical Chemistry Analyzer might assign a test result to an incorrect sample ID under the following workflow conditions. After loading a barcoded sample tube (sample A) onto the DxC 500 AU Clinical Chemistry Analyzer, all tests ordered for that barcoded sample tube are locked into the specific position of that barcoded tube on the analyzer (for example, the position on the DxLab rack or position in the stat table) until all tests are completed or cancelled for that sample tube. A problem might occur if, for any reason, all tests ordered for that barcoded sample tube (sample A) are not completed or remain "in pending", and the operator attempts to analyze a second sample tube (sample B) in that same position on the rack or stat table. If the bar code for sample B is not successfully scanned by the analyzer or if the sample tube is not labelled, the analyzer assumes the sample that previously occupied that position has been reloaded (sample A). The analyzer will complete and report the pending tests associated with sample A, using sample B. No results will be reported for sample A. This issue does not arise if all barcoded sample tubes are successfully scanned or if testing is complete for the first barcoded sample (sample A). Recall Start Date: March 7, 2025

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime