Health Products

ABL90 Flex and Flex Plus

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: ABL90 Flex and Flex Plus
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
ABL90 Flex PlusRWIN10 version 1.1 (933-894)
RWIN10 version 1.0 (933-803)
393-092
ABL90 FlexRWIN10 version 1.1 (933-894)
RWIN10 version 1.0 (933-803)
393-090

Issue

There is a potential risk with ABL90 Flex & ABL90 Flex Plus running RWIN10 1.0 and RWIN10 1.1. The issue is that the analyzer software may unexpectedly freeze during a sample measurement. The analyzer software includes a watchdog that detects a freeze and automatically restarts the software. If the analyzer freezes, a sample being processed will be lost. Recall Start Date: March 11, 2025

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime