Health Products

Stryker Instruments System 8 and 9 Saws

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Dental, Health products - Medical devices - Orthopaedics, Health products - Medical devices - General and plastic surgery
Product: Stryker Instruments System 8 and 9 Saws
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
System 8 Reciprocating Saw24096248238206-000-000
System 9 Sternum SawNot applicable.9207-000-000
System 9 Reciprocating SawMore than 10 numbers, contact manufacturer.9206-000-000

Issue

There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025