· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Epoc Blood Analysis Bgem Test Card | Lot with prefix 02 or 12 | 10736515 |
| Epoc Blood Analysis System - Single Use Blood Gas, Electrolyte And Metabolite Test Card | 10736382, Lot prefix 00 or 06 | CT-1006-00-00 |
Siemens confirmed that the occurrence of discrepant high pH results in samples introduced with higher injection volumes (beyond the epoc reader beep or after the "analyzing sample" message is displayed) with epoc sensor configuration 45.n (software version: epoc host v3.41.2 and epoc nxs v4.14.9 & 4.14.11) Recall start date: March 25, 2025