· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Carestation 650 | More than 10 numbers, contact manufacturer. | 1012-9650-000 |
| Carestation 620 | SM625030002MA SM623500001MA SM624280002MA SM623500002MA SM624370006MA SM623140017MA SM624320007MA SM624370005MA SM625040002MA SM623280016MA | 1012-9620-000 |
| Carestation 750 A1 | SQS23200001MA SQS24190110MA SQS24290021MA | 1012-9750-000 |
GE Healthcare has become aware that certain Carestation 620/650/650c and 750/750c anesthesia delivery systems (see affected products list) will not provide effective ventilation in volume control ventilation (VCV) mode. In these systems, effective ventilation can be achieved in pressure control ventilation (PCV) or pressure control ventilation volume guarantee (PCV-VG) modes or with manual ventilation. Recall start date: April 1 2025