· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| CADD-Solis Ambulatory Infusion Pump | More than 10 numbers, contact manufacturer. | 21-2101-0100-95 21-2101-0200-00 21-2101-0200-02 21-2111-0400-78 21-2111-0403-78 21-2112-0200-51 21-2112-0400-51 21-2112-0401-51 21-2112-0403-78 21-2101-0200-232 21-2102-0200-249 21-2111-0200-249 21-2112-0100-00 21-2112-0400-249 21-2101-0200-95 21-2102-51 21-2111-0100-51 21-2111-0300-95 21-2111-0401-78 21-2111-0403-92 21-2112-0300-95 21-2112-0401-78 21-2112-0403-92 21-2101-0200-249 21-2102-0200-95 21-2111-0300-00 21-2111-0400-51 21-2111-0401-51 21-2111-0403-51 21-2112-0200-02 21-2112-0300-50 21-2112-0400-78 21-2112-0403-02 21-2112-0403-51 21-2111-0200-232 21-2111-0300-02 21-2111-0400-249 21-2112-0200-00 21-2112-0200-95 21-2112-0300-02 21-2102-0200-00 21-2111-0300-50 21-2111-0400-01 21-2112-0100-51 21-2111-0100-249 2110 |
CADD-Solis pumps may lose communication with the CADD wireless communication module if wireless network setting changes on the hospital network are not compatible with the CADD wireless communication module. This may result in a "wireless module intermittent connection" high priority alarm, which will stop an ongoing infusion. Recall Start Date: April 8, 2025