Hugo RAS Tower, 120v

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Hugo RAS Tower, 120v
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Hugo RAS Tower, 120v10884521826649MRASC0003

Issue

The loss of motion has been found to occur intraoperatively due to an instrument failure, causing the system to halt teleoperation with the robotic instrument. While the instrument is disabled and during the user's removal of the instrument it is possible for the user to trigger an error code by applying excessive torque which also disables the arm cart assembly (ACA). In both scenarios (disabled instrument, and disabled ACA) halting of teleoperation is an intended response of the system to prevent further use of the robotic instrument or arm cart assembly until the issue is resolved, or the portion of the system with issue is replaced. Recall Start Date: April 29, 2025