Health Products

Midas Rex Legend, Burr-Tapered

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Midas Rex Legend, Burr-Tapered
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Midas Rex Legend, Burr-Tapered0228714562
0228582219
0228582216
0228714563
F2/8TA23S

Issue

In October 2024, Medtronic became aware of a manufacturing issue involving Midas Rex Legend Fluted Spiral Tool that rendered the tools unusable. The cutting head of the F2/8TA23S tool is manufactured on a CNC grinder from a rod of material. The grinding process can leave excess material (burrs) as it grinds the cutting flutes into the stem. A deburring process downstream removes this material. It was determined that the process was not adequately performed for approximately six months in 2024. These burs can prevent the tool from being inserted into an attachment and therefore prevent the device from being used to perform a surgical procedure. Recall Start Date: April 30, 2025