Health ProductsType I recall

Adult Manual Resuscitator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Adult Manual Resuscitator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Adult Manual Resuscitator526796
526804
526797
526805
526817
526782
526802
526815
526807
526811
526800
526806
526809
526814
526816
526798
526808
526790
526810
526818
CPRM1116F

Issue

Medline was informed by its contract manufacturer that the Adult Manual Resuscitator with mask, bag reservoir, and filter, CPRM1116F, was misassembled. Instead of the filter pre-attached at the exhalation port, it was connected to the patient port where the mask attaches. If the filter is incorrectly attached and not noticed before patient use, there could be a potential delay in patient resuscitation, which may result in life-threatening consequences, including hypoxia, hypercapnia, organ failure or death. Recall Start Date: May 9, 2025