Health ProductsType I recall

Newport HT70 Ventilator Plus

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Newport HT70 Ventilator Plus
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Newport HT70 Ventilator PlusHT7P190119
N16HT720719595
N16HT720719596
HT70PM

Issue

The C92 capacitor of HT70 ventilators may fail, disabling the output of the power management chip causing the ventilator to cease operation and stop ventilation. An audible alarm will be initiated, and the safety valve will open. In addition to the C92 issue, HT70 ventilators containing C159 capacitors reported failures of the shutdown buzzer alarm test, and as a result, the alarm "hardly sounded" or was of insufficient volume or duration per specification, placing a patient at risk of respiratory failure due to unexpected loss of ventilation. Recall start date: May 21, 2025