· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Video Gastroscope Therapeutic Type; Working Length 1100mm | 8G411K033 | EG-760CT |
| 3D Navigation Colonoscope | 1C754K057 | EC-760S-G/L |
| Led Adult HD Colonoscope, ELUXEO HD Gastroscope, Slim Optical Zoom Endoscope Diameter 11.8 | 2C731K155 KG402K232 8C729K097 JC729K932 7C729K100 7G402K360 JG402K452 5C729K533 5C729K534 | EC-760ZP-V/L EG-760R EC-760R-V/L |
| 720 Adult Colonoscope | 1C742K226 | EC-720R/L |
The devices were recently repaired at Vantage's service centre and returned to the customers. It came to their attention that the irrigation water pressure of these device(s) may exceed the specified range, potentially resulting in stronger-than-expected pressure during a procedure. Recall start date: May 5, 2025