Health Products

CareLink SmartSync™ Applications

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use, Health products - Medical devices - Cardiovascular
Product: CareLink SmartSync™ Applications
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
CareLink SmartSync™ Device Manager ApplicationAll lots.CARELINK SMARTSYNC DEVICE MANAGER APP
CareLink SmartSync™ Azure™ Astra™ ApplicationAll lots.D00U003
CareLink SmartSync™ Cobalt™ Crome™ ApplicationAll lots.D00U005
CareLink SmartSync™ Evera™ MRI ApplicationAll lots.D00U010
CareLink SmartSync™ Micra™ AV ApplicationAll lots.D00U007
CareLink SmartSync™ Viva™ Consulta™ Syncra™ Advisa™ Ensura™ ApplicationAll lots.D00U008
CareLink SmartSync™ PSA ApplicationAll lots.D00U002
CareLink SmartSync™ VisiaAF™ ApplicationAll lots.D00U011
CareLink SmartSync™ Viva™ Brava™ Evera™ ApplicationAll lots.D00U012
CareLink SmartSync™ Micra™ VR ApplicationAll lots.D00U006
CareLink SmartSync™ Micra™ VR2 ApplicationAll lots.D00U022
CareLink SmartSync™ Micra™ VR2 ApplicationAll lots.D00U009
CareLink SmartSync™ Micra™ AV2 ApplicationAll lots.D00U022
CareLink SmartSync™ Percepta™ Serena™ Solara™ ApplicationAll lots.D00U004

Issue

This recall pertains to a device electrical reset message being displayed inappropriately (i.e., when a true device electrical reset did not occur). There are two issues when the device electrical reset message is displayed inappropriately: Issue #1: The user is informed there was a device electrical reset when there was not. This can cause confusion and might lead the user to believe the device is no longer functioning properly. Issue #2: One of the messages that can be displayed inappropriately recommends device explant ("serious device error: device should be explanted"). While the message may be factually correct for the type of electrical reset it is designed to indicate, if displayed inappropriately it increases the risk of explant and patient harm. Recall start date: June 18, 2025

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime