Beckman Coulter Au Immunochemi Sys- Unsaturated Iron Binding Capacity (UIBC)
All lots.
OSR61205
Issue
UIBC OSR61205 is not meeting interference claims for haemoglobin interference serum application. As per product ifus the claim is less than 10% up to 2g/l haemoglobin. While the high pool is meeting the claim, the low pool is failing the hemolysate interference specifications with results of -43.63% or -8.95 μmol/l. Recall start date: July 1, 2025