· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Color Cuff - Sterile Disposable Tourniquet Cuffs | More than 10 numbers, contact manufacturer. | 5921-018-235 |
| Color Cuff - Sterile Disposable Tourniquet Cuffs | More than 10 numbers, contact manufacturer. | 5921-018-135 |
| Color Cuff - Sterile Disposable Tourniquet Cuffs | More than 10 numbers, contact manufacturer. | 5921-024-235 |
There is potential for the tourniquet cuffs flange to detach from the bladder where the welding connection is, which may result in the cuff not able to maintain air pressure. On February 12, 2025, a nonconformance was opened to address a manufacturing nonconformance identified through a Stryker complaint investigation for the tourniquet cuff flange becoming detached from the bladder. Less noticeable detached flanges have been reported and events confirmed. Recall start date: June 27, 2025