StrykeFlow Suction/Irrigator Jul 16, 2025 · Published by Health Canada · Official notice
Share
Alert type : Health product recall
Category : Health products - Medical devices - Gastroenterology and urology
Product : StrykeFlow Suction/Irrigator
Issue : Medical devices - Performance
What to do : Contact the manufacturer if you require additional information.
Audience : Healthcare
Recall class : Type II
We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.
Get health product recall alerts CA$3.99/mo · cancel anytime
Affected products Lot or serial number Model or catalog number StrykeFlow Suction/Irrigator More than 10 numbers, contact manufacturer. 0250070500 StrykeFlow Suction/Irrigator More than 10 numbers, contact manufacturer. 0250070520
It has been identified that due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the StrykeFlow 2 Suction Irrigator. Recall start date: July 2, 2025