Health Products

Osstell® Smartpeg Type 101-104

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Dental
Product: Osstell® Smartpeg Type 101-104
Issue: Medical devices - Performance
What to do: Contact the manufacturer Osstell has a new IFU under implementation and has initiated a design change.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Osstell® Smartpeg Type 101-104Not applicable.100766
Osstell® Smartpeg Type 101-104Not applicable.100767
Osstell® Smartpeg Type 101-104Not applicable.100769
Osstell® Smartpeg Type 101-104Not applicable.100768

Issue

If not correctly used with a reference measurement on implant level. The measurement on abutment can indicate a false high ISQ value. This can influence the clinician to take the wrong decision and load the implant too early, in the extension potentially resulting in implant failure for the patient. Recall start date: May 23, 2025